DIN 6800-2:2026-11 specifies a dose measurement method using ionization chambers for high-energy photon and electron radiation in teletherapy. It focuses on determining absorbed dose to water and absorbed dose rate to water using vented, air-filled ionization chambers. The standard covers, among other things, ⁶⁰Co gamma radiation and photon and electron radiation from clinical linear accelerators, providing a consistent basis for precise dosimetric measurements.
The standard describes ionization chamber dosimetry in high-energy photon and electron radiation fields. It specifies requirements for the selection and positioning of ionization chambers, water phantoms, reference conditions, calibration coefficients and correction factors. It covers measurements under reference and selected non-reference conditions, as well as the determination of measurement uncertainty.
The scope includes photon radiation from clinical linear accelerators with and without flattening filters, as well as electron radiation from 4 MeV to 25 MeV. Dose measurements on MR-linacs and in FLASH radiation fields are not covered by this part of the standard.
Precise dosimetry is essential for the quality-assured use of high-energy radiation in radiotherapy. DIN 6800-2 supports medical physics and dosimetry professionals in reproducibly determining absorbed doses to water under defined conditions and evaluating measurement uncertainties transparently.
The updated requirements for chamber positioning, correction factors and the calibration of ionization chambers are particularly relevant in practice. They take account of modern linear accelerators and current findings in clinical dosimetry.
The 2026 edition has undergone a comprehensive technical revision. Key changes include:
DIN 6800-2:2026-11 provides a key basis for ionization chamber dosimetry of high-energy photon and electron radiation. The new edition takes account of current dosimetric findings and updates essential requirements for measurement and calibration in radiotherapy.
This document replaces DIN 6800-2:2020-08 .